Gas Anaylzers

Description

Equipment (DONE.) Quiz on Gas Anaylzers, created by LeeAnna Shepherd on 12/08/2016.
LeeAnna Shepherd
Quiz by LeeAnna Shepherd, updated more than 1 year ago
LeeAnna Shepherd
Created by LeeAnna Shepherd over 8 years ago
22
0

Resource summary

Question 1

Question
Gas Chromatography (Absorption Type) Sample gas moves through a [blank_start]separator[blank_end] column where the differing sizes of molecules for the individual gases make each gas exit at a different [blank_start]rate[blank_end],
Answer
  • separator
  • rate

Question 2

Question
Advantages: Analyzes [blank_start]many gases[blank_end] at once. Very accurate Disadvantages: Requires more [blank_start]time[blank_end] than other analyzers. Affected by: [blank_start]H2O[blank_end] and [blank_start]CO2[blank_end] Measures: [blank_start]Ne[blank_end], [blank_start]O2[blank_end], [blank_start]N2[blank_end], [blank_start]CO[blank_end]
Answer
  • CO
  • N2
  • O2
  • Ne
  • CO2
  • H2O
  • time
  • many gases

Question 3

Question
Mass Spectometry Function: Sample is drawn in by vacuum to the [blank_start]ionization[blank_end] chamber. Individual gases are [blank_start]separated[blank_end] and analyzed. Measures: Gas percentages Advantages: [blank_start]Rapid[blank_end] Response time, breath by breath analysis, multiple gas analysis Can analyze: [blank_start]He[blank_end], [blank_start]N2[blank_end], [blank_start]O2[blank_end], [blank_start]CO2[blank_end] Disadvantages: Expensive, large, high maintenance Most Suitable for: analysis of [blank_start]several[blank_end] patient at once rather than individual testing.
Answer
  • ionization
  • separated
  • Rapid
  • He
  • N2
  • O2
  • CO2
  • several

Question 4

Question
CO2/CO/CH4 AKA: Infrared Absorption Analyzer Quick response time is suitable for [blank_start]DLCO[blank_end] Calibrate: to room air ([blank_start]0[blank_end]%) and known concentration ([blank_start].3[blank_end]%) CO then re-zero and use [blank_start]gain[blank_end] to match read out to [blank_start]known[blank_end] concentration. Changes in altitude do [blank_start]NOT[blank_end] affect readings.
Answer
  • DLCO
  • 0
  • gain
  • known
  • NOT
  • .3

Question 5

Question
Helium analyzer Also Known as: [blank_start]Thermal[blank_end] Conductivity or [blank_start]Wheatstone[blank_end] Bridge Measures: Gas [blank_start]percentage[blank_end] Calibration: To room air ([blank_start]0[blank_end]%) and known ([blank_start]10[blank_end]%) then use gain to adjust to [blank_start]known[blank_end] amount. Do not use with [blank_start]flammable[blank_end] gases. Used with Helium dilution tests.
Answer
  • Thermal
  • Wheatstone
  • percentage
  • 0
  • 10
  • known
  • flammable

Question 6

Question
Nitrogen analyzer Also known as: [blank_start]Geissler[blank_end] Tube Ionizer Sample is pumped into [blank_start]ionization[blank_end] chamber where light is emitted and monitored by a phototube. Intensity of light is directly [blank_start]proportional[blank_end] to N2% Measures gas percentage Capable of breath by breath analysis Calibration: To zero with [blank_start]100[blank_end]% O2 and a [blank_start]known[blank_end] concentration.
Answer
  • Geissler
  • ionization
  • proportional
  • 100
  • known

Question 7

Question
O2 Analyzers Galvanic [blank_start]Fuel[blank_end] Cell Analyzer: Measures partial [blank_start]pressure[blank_end] and displays [blank_start]FiO2[blank_end] as %. If unable to calibrate you should replace the [blank_start]fuel[blank_end] cell. Calibration during exercise testing: Low [blank_start]12[blank_end]-[blank_start]14[blank_end]% and high [blank_start]21[blank_end]% Poloragraphic Analyzer Similar to galvanic but uses a [blank_start]battery[blank_end] Analogous to the [blank_start]Clark[blank_end] electrode Calibration is to [blank_start]21[blank_end]% and [blank_start]100[blank_end]%
Answer
  • Fuel
  • pressure
  • FiO2
  • fuel
  • 12
  • 14
  • 21
  • 21
  • 100
  • Clark
  • battery

Question 8

Question
Gas & Water Absorbers CO2 Absorber: AKA Sodium [blank_start]Hydroxide[blank_end], NaOH (soda [blank_start]lime[blank_end] or soda-sorb), Baralyme Required for [blank_start]rebreathing[blank_end] tests (FRC by He Dilution, DLCO-RB) Replace scrubber: When [blank_start]exhausted[blank_end], when [blank_start]color[blank_end] changes, or weekly. CO2 Scrubber produce [blank_start]water[blank_end] as a by product and should be placed [blank_start]upstream[blank_end] from water absorbers.
Answer
  • lime
  • Hydroxide
  • rebreathing
  • exhausted
  • color
  • water
  • upstream

Question 9

Question
Gas & Water Absorbers Water Absorber ([blank_start]Dessicant[blank_end]) Drierite ([blank_start]Calcium[blank_end] Sulfate) Removes moisture as patient breathes through system. Proper placement of absorbers: 1st: [blank_start]CO2[blank_end] Absorber 2nd: [blank_start]H2O[blank_end] Absrober 3rd: Anaylzer Permapure (fafion Tubing) Conditions gas to [blank_start]equilibrate[blank_end] with ambient humidity Allows for accurate correction for humidity in a gas sample.
Answer
  • Calcium
  • Dessicant
  • CO2
  • H2O
  • equilibrate
Show full summary Hide full summary

Similar

Exchange Surfaces and Breathing Quiz
gordonbrad
Particle Model of Matter
Brodie McMeowface
States Of Matter
Paige Louise
Chromatography
Tom Johnson
Density Unit 2.4
Assged Abdalla
Laboratory Tools and Equipment
Mark Anthony Sy
Matter, Physical and Chemical Changes Quiz
Jeffrey Piggott
Identify common tools and equipment
T Andrews
states of matter
fabiana monge
New Weapons
izzykersley01
P&ID
Thái Vũ